4.2
SPAN CALIBRATION
4.2.1
Adjust ZERO POT to a 000 display count as
suggested in section 4.1.
4.2.2
Apply full scale signal per given strain gage. For
3mV/V and 1.24 V excitation, apply 3.720 mV at
± SIGNAL input.
4.2.3
Use SPAN POT to fine tune to a 1999 display count.
4.3
After proceeding through the steps above, apply a
ZERO signal again and check for a 000 display count.
If the ZERO has shifted, repeat the steps for ZERO
and SPAN calibration.
JUMPER SWITCH
V
U
T
S
R
Q
SOLDER SWITCH
A
B
C
C
ALIBRATION
Table 4.4

S3 JUMPER SWITCH DESCRIPTION

SELECTED DECIMAL POINT POSITION
Table 4.5

SOLDER SWITCH DESCRIPTION

Output is connected to SPAN POT. See NOTE 4.
Gain Resistor R6 (750 Ohm) is selected.
Gain Resistor R7 (OPEN) is selected.
& G
S
AIN
ELECTION
FUNCTION
TEST SEGMENT
DUMMY ZERO
1XXX.0
1XX.X0
1X.XX0
1.XXX0
FUNCTION
See Table 4.1 and 4.2.
Default OPEN. See NOTE 5.
4
13