Edit Or Delete Qal3 Values; Exporting Or Deleting Qal3 Data - ABB EasyLine EL3000 Operating Instruction

Continuous gas analyzers
Hide thumbs Also See for EasyLine EL3000:
Table of Contents

Advertisement

EL3000 CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D

Edit or delete QAL3 values

Edit a single QAL3 value
1. On the 'QAL3 Monitoring' page in the display of the control
chart (see Display and print control chart on page 162), click
on the number of the QAL3 value (column 'No.') that needs to
be edited.
2. The following information and values can be changed on the
'Edit Data' page:
'Technician' (technician),
'Comment' (comment),
'sAMS' (s
-value for zero point or reference point).
AMS
3. Leave the page by clicking 'Save' (save) or 'Cancel' (cancel).
Deleting a single QAL3 value
1. On the 'QAL3 Monitoring' page in the display of the control
chart (see Display and print control chart on page 162), click
on the number of the QAL3 value (column 'No.') that needs to
be deleted.
2. On the 'Edit Data' page, click on 'Delete' (delete).
3. Confirm the following question (Do you really want to delete
this value?) by clicking on 'OK'.
4. The value is deleted and the control chart is displayed again.

Exporting or deleting QAL3 data

Exporting QAL3 data
The QAL3 data can be exported from the gas analyzer to archive
it or to for example edit it in a spreadsheet program.
1. In the QAL3 monitoring web interface, click 'QAL3
Monitoring'.
2. Select the type of display on the 'QAL3 Monitoring' page:
'Zero point simple' (simple zero point),
'Reference point simple' (simple reference point),
'Zero point details' (detailed zero point),
'Reference point details' (detailed reference point).
3. Select the type of control chart (see 'Control cards on
page 160') ('Control Chart to Display') for the desired sample
component.
4. At the bottom of the page, click on 'Export data' (Export
data).
The QAL3 data from the displayed control chart is exported
to a text file (.txt).
5. Open the text file or save it under a new name as needed.
6. If necessary, repeat steps 2 to 5 for additional control charts.
Deleting QAL3 data
1. In the QAL3 monitoring web interface, click 'QAL3
Monitoring'.
2. On the 'QAL3 Monitoring' page, click on 'Settings' (settings).
3. In the 'Data storage' window (save data), click on 'Delete all
data' (Delete all data).
4. A warning appears:
'Do you really want to delete all the data? This operation
cannot be undone.
Make sure all data has been exported before proceeding.'
5. Click on 'OK'. The QAL3 data is deleted.
6. Leave the page by clicking 'Save' (save) or 'Cancel' (cancel).