ABB EasyLine EL3000 Operating Instruction page 122

Continuous gas analyzers
Hide thumbs Also See for EasyLine EL3000:
Table of Contents

Advertisement

122
EL3000 CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
... 9 Calibration
... Automatic calibration: settings
Calibration method (automatic)
Analyzer
Calibration method
Uras26
Zero point / end-point calibration
Limas23
Zero point / end-point calibration
Magnos206
Zero point / end-point calibration or substitute gas
calibration or single-point calibration
Magnos28
Zero point / end-point calibration or substitute gas
calibration or single-point calibration
Magnos27
Zero point / end-point calibration or substitute gas
calibration
Caldos27
Zero point / end-point calibration or single-point
calibration
Fidas24
Substitute gas calibration
Oxygen sensor
Single-point calibration
Remark
Cannot be changed.
Cannot be changed.
If a substitute gas component is set up in the Magnos206 analyzer, the calibration
method is factory-set to substitute gas calibration, otherwise the zero point / end-
point calibration is set. Alternatively, single-point calibration can be selected.
See also page 130.
If a substitute gas component is set up in the Magnos28 analyzer, the calibration
method is factory-set to substitute gas calibration, otherwise the zero point / end-
point calibration is set. Alternatively, single-point calibration can be selected.
See also page 131.
If a substitute gas component is set up in the Magnos27 analyzer, the calibration
method is factory-set to substitute gas calibration, otherwise the zero point / end-
point calibration is set.
See also page 133.
With standard gas, this cannot be changed.
Cannot be changed.
Cannot be changed.