4.1.2 System test
TRM, Centiva/5 Plus, M1047448/ CM-922026, rev. 3.50.a of 11/05
After having passed the
power-up test, the
Centiva/5 Plus
ventilator prompts with a
short audible sound and
moves to the "Start
system test" screen:
Connect the Centiva/5
Plus ventilator to a
breathing circuit and
seal the open end of the
y-piece by pushing it
onto the sealer located on the expiratory valve block. Select the type of patient,
appropriate to the breathing circuit attached.
The system test parameter will vary with the patient type selected to match the
Note
breathing circuit's specifications.
Push the ComWheel to start the system test. This system test takes at least 40
secs. The remaining test time counts down in the screen.
During the system test, the system will check proper function of:
O
gas supply
2
O
flow valves
2
O
flow sensor
2
AIR gas supply
AIR flow valves
AIR flow sensor
Nebulizer gas
supply
Pressure
sensors
Expiration valve
Safety valves
Exp. flow sensor
O
sensor
2
Back up battery capacity
Centiva/5
+
Bi-Level VG
Newborn
Trend
SW 3.nn.p HW 0/r
System test
Complete system set up, seal y-piece
Start system test by pressing
Ensure silver exp. valve is in place
Body weigth
55 kg
Adult
Pediatric
Newborn
Centiva/5 +
Bi-Level VG
Newborn
Trend
SW 3.nn.p HW 0/r
System test
Test active, wait for 40 secs.
Adult
4 Test and calibration
09:36:53
2005-03-21
Bypass
system test
Unlock
exp. valve
<Enter>
Calibrate
O2 sensor
Align
touch screen
Retain
trend data
Enter
4-3