This chapter draws attention to important safety precautions and introduces the PathSpeed CR SP1001. PathSpeed CR SP1001features Safety precautions Safety compliance Operating modes The user interface Switching on the PathSpeed CR SP1001 Switchin off the PathSpeed CR SP1001 OI10026 PathSpeed CR User Manual...
PathSpeed CR SP1001 features The PathSpeed CR SP1001 is a digitizer for image plates retain- ing latent X-ray images. The PathSpeed CR SP1001 accepts one cassette containing one image plate at a time. The PathSpeed CR SP1001: • takes the cassette containing the image plate from the cassette slot;...
As is the case for all technical devices, the PathSpeed CR SP1001 must be operated, cared for and serviced correctly. • If you don’t operate the PathSpeed CR SP1001 correctly or if you don’t have it serviced correctly, GE is not liable for resulting distur- bances, damages or injuries.
Page 10
• If you notice conspicuous noise or smoke, disconnect the Path- Speed CR SP1001 immediately. • Check that the voltage setting of the machine matches the power supply voltage before connecting the machine to the mains. OI10026 PathSpeed CR User Manual...
Page 11
Safety instructions for laser products The PathSpeed CR SP1001 is a Class 1 Laser Product. Under normal operating conditions, when the service doors are closed, there can be no laser radiation outside the PathSpeed CR SP1001. • Open the lower front door only to replace the erasure lamps or the fuses.
Operating modes The PathSpeed CR SP1001 can be operated in three modes: operator mode, key-operator mode, and service mode. Operator mode The operator mode groups all basic functions which are aimed at radiographers: • Reading an image plate; • Reading an emergency image plate;...
‘Survey of advanced func- tions (‘Key-operator mode’)’ on page 40. For detailed information on the key-operator mode, refer to the PathSpeed CR SP1001 Reference manual. The PathSpeed CR SP1001 interfaces with the user via: •...
Page 14
Status indicator Emergency buttons Keypad and display OI10026 PathSpeed CR User Manual...
Page 15
The keypad The PathSpeed CR SP1001 keypad features the following keys: To give an image the status ‘emergency’ when it is Emergency sent to the image processing device. This key can only be used for cassettes with ID data. To erase images without digitizing them.
Page 16
• To move the cursor to the previous entry field. Up key • To scroll upwards. • To increment the number in a numeric entry field. • To move the cursor to the next entry field. Down key • To scroll downwards. •...
Page 17
The display The PathSpeed CR SP1001 control panel has a backlit LCD dis- play with 8 lines of 40 characters each. Its lay-out depends on the operating mode. In operator mode, the display has dedicated areas for specific information: Set-up display...
Page 18
Extra comment or action to take System status: • READY: The PathSpeed CR SP1001 is ready for operation. • BUSY: The PathSpeed CR SP1001 is treating an image plate. • ERROR: An error has occurred. Refer to ‘Troubleshooting checklist’ on page 42.
Page 19
When the is treating an image plate, it dis- PathSpeed CR SP1001 plays the following screen: BUSY Miller Chest AP In key-operator mode, operation is menu driven. The menu displays the key-operator functions, the active keys, and the service code.
Emergency buttons Two emergency buttons are located at th the PathSpeed CR SP1001. They permit ing emergency image plates without ID d upper button permits processing of em images of the trunk; the lower button per...
Page 21
The length of the beep indicates the response of the system to a key command. • A short beep means that PathSpeed CR SP1001 has accepted the key command and is starting the operation. • A long beep means that you have pressed a non-active key or that the PathSpeed CR SP1001 has rejected the key command.
Switching on the PathSpeed CR SP1001 Make sure that the setting of the voltage selector at the back of the machine matches the power supply voltage. 230-240 Locate the main switch and place it in position ‘I’. OI10026 PathSpeed CR User Manual...
Page 23
The following screen is displayed: WAIT Self test proceeding During the self-test, you cannot activate any functions. If the PathSpeed CR SP1001 has completed the self-test suc- cessfully, the PathSpeed CR SP1001 enters the operator mode and displays the operator main screen: READY...
It is recommended to switch off the PathSpeed CR SP1001 at the end of the day. Only switch off the PathSpeed CR SP1001 if you do not intend to digitize emergency image plates overnight. Switching on the PathSpeed CR SP1001 takes approximately 3 minutes.
Page 25
Chapter Basic operation (‘Operator mode’) This chapter provides basic information on how to digitize image plates under normal conditions and in emergency situations. It also treats how to re-erase an image plate to prevent ghost images caused by previous exposures or by stray radiation.
PathSpeed CR SP1001 must display the operator main screen with ‘Ready’ status, e.g.: VIPS not ready READY Status field PathSpeed CR SP1001 • the status indicator at the top of the PathSpeed CR SP1001 must be green and be lit constantly. OI10026 PathSpeed CR User Manual...
Page 27
PathSpeed CR SP1001 as shown below. Make sure to insert the cassette with the hinge [1] at the top and the locking mechanism [2] at the bottom. While treating the image plate, the PathSpeed CR SP1001 will dis- play the following screen: BUSY...
Page 28
When the PathSpeed CR SP1001 returns the cassette, it is ready to be re- used immediately. However, if you leave it for more than 3 days before re- using it, you must re-erase it first. Refer to ‘Re-erasing an image plate’ on...
Status field ADC SOLO • the status indicator at the top of the PathSpeed CR SP1001 must be green and be lit constantly. The PathSpeed CR SP1001 is operational if the status field equals ‘READY’, even if status messages of the destination are shown (e.g.
Page 30
Next cassette gets emergency status ADC SOLO The emergency status will only be assigned to the first image pl you insert into the PathSpeed CR SP1001 cassette slot after pre Emergency key. Insert the cassette containing the em image plate into the cassette slot as show...
Page 31
function and return to the operator main screen. OI10026 PathSpeed CR User Manual...
Page 32
‘Ready’ status, e.g.: Status READY ADC SOLO • the status indicator at the top of the PathSpeed CR SP1001 must be green and be lit constantly. Press one of the emergency buttons at the front of the PathSpeed CR SP1001 For digitizing unidentified emergency images of the trunk.
Page 33
[2] at the bottom. The image plate will be digitized using the scan parameters for the emergency button as defined during configuration. When the PathSpeed CR SP1001 has treated the emergency image digital image data plate, it displays the operator main screen. The are transmitted to the image processing device accompanied by default ID data.
PathSpeed CR SP1001 must display the operator main screen with ‘Ready’ status, e.g.: Status READY ADC SOLO • the status indicator at the top of the PathSpeed CR SP1001 must be green and be lit constantly. OI10026 PathSpeed CR User Manual...
Page 35
Put cassette in slot or press to quit Insert the cassette into the cassette slot. While erasing, the PathSpeed CR SP1001 will still display the above screen. When the PathSpeed CR SP1001 has erased the image plate, it displays the operator main screen. Warning...
Page 36
While erasing, the PathSpeed CR SP1001 will display: READY ERASURE WARNING The next cassette will be erased Put cassette in slot or press to quit When the PathSpeed CR SP1001 has erased the image plate, it displays the operator main screen. OI10026 PathSpeed CR User Manual...
Page 37
Status READY ADC SOLO • the status indicator at the top of the PathSpeed CR SP1001 must be green and be lit constantly. Insert the cassette into the cassette slot. The PathSpeed CR SP1001 will automatically erase the image plate.
Page 39
Chapter Advanced operation (‘Key-operator mode’) This chapter gives an overview of the key-opera- tor functions, preventive maintenance actions and troubleshooting. For detailed information on these topics, refer to the Reference Manual. Survey of advanced functions (‘Key-oper- ator mode’) Preventive maintenance work Troubleshooting checklist OI10026 PathSpeed CR User Manual...
Function in Section in Reference Manual key-operator main menu Queue management Consulting the image transmission queue (‘Queue management’). Customizing the PathSpeed CR SP1001 Digitizer set-up (‘Digitizer set-up’). Setting the date and time. Date and Time Send test image Sending test images.
Preventive maintenance work The only maintenance action which you must perform is checking the image quality. Refer to the Reference manual of the image processing system. OI10026 PathSpeed CR User Manual...
A survey of possible problems is listed below. If corrective actions are straightforward, they are given below. The more elab- orate troubleshooting procedures are explained in detail in Chap- ter 4, ‘Preventive maintenance and troubleshooting’ of the PathSpeed CR SP1001 Reference manual. General errors Error Action PathSpeed CR SP1001 does Refer to ‘Checking the voltage supply’...
Page 43
Status field: ERROR Error field: ‘CODE XXXXX ’ MESSAGE 1 Message 2 Action 1. Check setting of voltage selector switch on back Refer to ‘Checking the voltage panel. POWER SUPPLY OUT supply’ on page 81 of the OF TOLERANCE 2. Check fuses of Reference manual.
Page 44
Status field: WARNING MESSAGE 1 Message 2 Action • Press Confirm key. • Refer to ‘Consulting the UNKNOWN Please press image transmission queue DESTINATION check queue (‘Queue management’)’ on [PPNAME] page 32 of the Reference manual. • Press Confirm key. English will be used.
Status field: LOCKED MESSAGE 1 Message 2 Action • Refer to ‘Consulting the image transmission queue (‘Queue management’)’ on page 32 of the Reference manual. IMAGE-QUEUE FULL Check queue • Check that the PathSpeed CR SP1001 is not off line (Refer to ‘The display’...
Specifications Product description Type of product Digitizer Commercial name ADC Solo Model number 5155 Original seller/manufacturer Agfa-Gevaert NV-Mortsel Labelling 93/42 EEC ‘Medical Devices’ (Europe) UL 1950, CSA 22.2 No. 950 (North America) (North America) Dimensions Length, at cassette slot 730 mm Length, at foot 700 mm Width...
Page 51
Power consumption Standby • 230 V/ 50 Hz configuration 230 W • 208 V/ 60 Hz configuration 227 W • USA: 120 V/ 60 Hz configuration 216 W • Japan: 100 V/ 60 Hz 220 W During operation • 230 V/ 50 Hz configuration max.
Page 52
Heat emission • During scanning max. 1610 W • Standby 230 W 60 - 72 secs Cassette return time End of Life Estimated product life (if regularly serviced and maintained 7 yrs. according to GE instructions) OI10026 PathSpeed CR User Manual...