Connections

21.3 Important Notes: Connecting Auxiliary Equipment

CAUTION !
The leakage current of the entire system including any / all auxiliary
equipment must not exceed the limit values as per EN60 601-1-1:1990 (IEC
60601-1-1) resp. other valid national or international standards. All
equipment must comply with UL, CSA and IEC requirements.
CAUTION !
Auxiliary equipment must only be connected to the main console with the
special mains outlet provided for the electrical safety of the system.
CAUTION !
Auxiliary equipment with direct mains connection requires galvanic
separation of the signal and / or control leads.
CAUTION:
The outlets ST1 – ST5 may be set to 230V or 115V
Modification of voltage setting only by an authorized service person!
ATTENTION !
With the color video monitor connected to its isolated mains outlet, the
remaining load capacity for auxiliary equipment is max. 350 VA.
WARNING !
Only accessories explicitly recognized by the system manufacturer
GE Medical Systems Kretztechnik GmbH & Co OHG may be used in
connection with the system.