8-16-2. Level Depandent Adjustment

8-16-3. Detail Frequency Adjustment

8-16-2. Level Depandent Adjustment
Preparation
. On the setting menu, set as follows.
PAGE
: FUNCTION 1/2
: DETAIL → ON
ITEM
: TEST OUT → ENC
ITEM
. OUTPUT/DCC switch (inside panel) → CAM/ON
. Shoot a gray-scale chart in the full underscan's picture
frame.
Setting point
: 1 Lens IRIS
: A = 100 ± 2 IRE (NTSC)
Spec.
A = 700 ±14 mV (PAL)
Adjustment procedure
Test point: TEST OUT connector
1. On the setting menu, adjust as follows.
PAGE
: LEVEL 1
ITEM
: LEVEL DEPEND
Spec.
: Eliminate the detail signal from portion
B.
Note
. After this adjustment, be sure to perform 8-16-4. H/V
Ratio Adjustment, and 8-16-5. Detail Level Adjustment,
in that order.
8-18
8-16-3. Detail Frequency Adjustment
Note
. Perform this adjustment, if necessary, to suit the custom-
er's preferences.
Preparation
. On the setting menu, set as follows.
PAGE
ITEM
ITEM
. OUTPUT/DCC switch (inside panel) → CAM/ON
. Shoot a gray-scale chart in the full underscan's picture
frame.
Setting point
Spec.
A
. Select the line at the center white portion of the gray-
scale chart.
Adjustment procedure
Test point: TEST OUT connector
1. On the setting menu, adjust as follows.
B
Note
. After this adjustment, be sure to perform 8-16-4. H/V
Ratio Adjustment, and 8-16-5. Detail Level Adjustment,
in that order.
: FUNCTION 1/2
: DETAIL → ON
: TEST OUT → ENC
: 1 Lens IRIS
: A = 100 ± 2 IRE (NTSC)
A = 700 ±14 mV (PAL)
PAGE
: LEVEL 1
ITEM
: DTL FREQ. (Factory setting: 0)
Spec.
: Set the desired width at the edge of
portion B.
A
B
DNV-5
DNW-7/90/90WS