1.5.4 System Menu

12
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
The System menu, pictured above, accesses the options that modify
the basic configuration of the PRO Monitor, test the EEPROM, and
print the error log.
Language Button
This button displays the language choices. The language dialog
boxes are not pictured. The PRO Monitor software can display menus
in six languages: UK, USA, German, French, Dutch, and Spanish.
Selecting any language button removes all other language buttons
from the screen, indicating that the remaining button is the chosen
language. Selecting Clear restores all the language buttons, allowing
the user to select again. OK saves the selection and a dialog box
requests that the Monitor be turned off. If no language is selected
when OK is pressed, the language is stored as undefined. In this
case, the Monitor prompts the user to select a language on every
power up until a selection occurs.
Comms Button
This button accesses the Communications menu. The settings on this
menu configure communications with a host computer when PRO
Monitors are connected to a network.
The Remote Op check box toggles remote operation. When on, a
check mark (√) appears in the box. When on, the PRO Monitor
responds to external commands and can initiate a BP determination
when prompted by a host computer. Remote operation requires
DINAMAP
Host Communications Protocol, which is described in the
®
DINAMAP
Host Communications Reference Manual.
®
The Standard check box sets the host comms protocol to standard
format, and the baud rate at 9600 bps.