Advertisement

...5 PROGRAMMING
...5.4 Configure Alarms

5.4.1 Wash Cycle Configuration (applicable only to Alarm 3)

A3: Type
set to Wash
Cont.
-----
Pulsed
Wash Mode
Hours
15
Mins
Wash Frequency
Mins
15
Secs
Wash Duration
1. 0
Mins
Recovery Period
Config. Alarm 3
Continuous
Pulsed
32
Wash Mode
The wash can be configured as continuous or pulsed. If
energized for the wash duration. If
second for the duration of the wash – see Fig. 5.1.
Wash Frequency
Wash frequency is set in 15 minute increments between 15 and 45 minutes, then in 1 hour
increments between 1 and 24 hours.
Wash Duration
Wash duration is set in 15 second intervals between 15 and 45 seconds, then in 1 minute
intervals between 1 and 10 minutes.
Recovery Period
The recovery period is set in 0.5 minute intervals between 0.5 and 5.0 minutes.
Configure Alarm 4.
Config. Alarm 4
See Section 5.5.
CONFIG. OUTPUTS
Wash duration
1s 1s
Fig. 5.1 Pulsed and Continuous Wash Cycles
is selected, the relay is switched on and off every
Pulsed
Frequency
Recovery period
is selected, the relay remains
Cont.
t
t