...5 PROGRAMMING

5.9 Test Outputs and Maintenance

-----
TEST/MAINTENANCE
-----
Test Outputs
4. 0 0
mA
20. 0
%
Test Output 1
-----
Maintenance
Auto.
On
-----
Off
Hold Outputs
Maintenance
38
Test Outputs
Displays the output test details for the four channels. Test Output 1 only is shown; the
remaining outputs are identical.
Note. Outputs 3 and 4 are available only if the optional analog output board is fitted.
See below.
Maintenance
Test Output 1
The theoretical output current value.
Output current as a percentage of the full range current.
Use the
and
keys to adjust the displayed theoretical output current value to give the
output required.
See Section 7.3.
FACTORY SETTINGS
Test remaining outputs.
Test Output 2
Maintenance
Hold Outputs
Enables the relay action and analog outputs to be maintained.
Auto.
– Hold is released automatically after six hours.
On
– Changes in relay action and analog outputs are inhibited.
Off
– Changes in relay action and analog outputs are not inhibited.
Note. The LEDs flash while the analyzer is in Hold mode.
Return to main menu.
TEST/MAINTENANCE
See Section 7.3.