WARNING:
1. Introduction
The XP Series is an easy to use, personal gas detector, designed for monitoring the atmosphere for
potentially hazardous levels of oxygen deficiency or toxic gas. It has built-in cell decay compensation,
thermal shock protection, and Reflex™, a patented sensor failure detection system, for maximum
reliability.
SAFETY INFORMATION
•
Substitution of components may impair intrinsic safety
•
Only use specified batteries: Energizer® CR2 or Sanyo CR2
•
Do not activate the detector after the date on the packaging
•
Perform a self-test prior to each day's use (See Section 6)
•
Do not use in oxygen-enriched atmospheres
•
Periodically test the sensor's response to gas by exposing the detector to a target gas concentration
that exceeds the alarm set points. Manually verify that the audible, visual and vibrating alarms
are activated.
•
Exposure to extremely high levels of over-range gas may cause temporary loss of sensor accuracy.
Allow detector to stabilize in a zero gas atmosphere before reuse. It is recommended to perform
a sensor response gas test prior to reuse.
IMPORTANT NOTICE
Honeywell Analytics can take no responsibility for the use of its equipment if it is not used in accordance
with the appropriate issues and/or amendments of the relevant manual. If further details are required
which do not appear in this manual, contact Honeywell Analytics or their agent.
Honeywell Analytics shall not be liable for any incidental or consequential damages in connection with
any deletions, errors or omissions in this manual.