Functional Checks
4.4.1 Save Small Backup (Scan Settings)
The Image/Scan Settings contain:
•
Application Settings
•
2D Factory and 2D User Presets
•
3D/4D Factory and 3D/4D User Presets
•
Annotation Presets
•
Scan Assistant Configuration
•
Measure Configuration
•
Biopsy Lines
1.
Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2.
Press the Utilities key on the control console.
3.
In the "Utilities" menu touch the
4.
On the left side of the screen select
5.
Click the
6.
Choose the media (2) and click the
7.
Select the
8.
Click the
4-30
System Setup button to invoke the setup desktop on the screen.
Backup and then click the System Configuration tab.
Save button (1) of the "Small Backup (Scan Settings)" group.
Save button (3).
New File... key and enter a file name (without extension).
OK key to start the process. When the saving has been completed, click OK .
Figure 4-22 Save window
Voluson E-Series Service Manual
KTD106657 Revision 2