8 Technical Specifications
8.1 400T System
8.1.1 Electrical characteristics
Mains voltage for the system
Heat dissipation
For further information, see the Tube technical data sheet, at the accompanying documents.
8.1.1.1 Classification
Classification according to IEC/EN 60601-1.
Class
Applied part
Protection against ingress of water
Mode of operation
Use of anesthetic mixtures
Classification according to IEC/EN 60601-1–2
Class
Operator´s Manual
2020-09-18
Technical Specifications
400 V 3N, 50/60 Hz
400 V 3~
480 V 3~
150 A (Short term peak value),
(required fuse 63 A thermal breaker, B-curve).
689 BTU/hr
Class I equipment. All dead metal parts of
the equipment are electrical connected to
protective earth.
Type B
IPX0
Intermittent operation: 20% 1 min ON / 4
min OFF
The equipment is not suitable for use in the
presence of flammable anesthetics mix-
tures with air, oxygen or nitrous oxide.
Class A
0180-095-418-EN Rev. 6.1
Page 133
400T System
OMNERA®