Uninterrupted Power Supply (UPS)

NOTE:
CAUTION
Ultrasound System
Direction 5444964-100 English Rev. 5
For all instructions in the "Electrical safety tests" section in case
of using a UPS (uninterrupted power supply) the terms outlet,
wall outlet, AC wall outlet and power outlet refer to the AC power
outlet of the UPS. In case of further available AC (or DC) power
outlets at the same used UPS, these must remain unused i.e.
not connected to any other devices.
Please observe that some Uninterruptible Power Supplies
(UPS) may not be medical devices! If the UPS is not a medical
device, it has to be located outside of the patient environment
(according to IEC 60601-1 / UL 60601-1).
1.5 m
1. Patient environment
Common Service Information
Electrical safety tests
2.5 m
1.5 m
4-15