Honeywell CIU 888 Firmware Upgrade Manual page 31

Hide thumbs Also See for CIU 888:
Table of Contents

Advertisement

FIGURE 2-10
FIGURE 2-11
Part No.: 4417596_Rev11
License Updates
1. Shut down every Entis Pro system connected to the CIU 888s in the
redundant system setup. For example, if in the system setup each
CIU 888 is connected to a separate Entis Pro system, both Entis Pro
systems need to be shut down.
2. Shut down the primary CIU 888:
a) Connect the PC/laptop to the Service port of the CIU 888 using a
Cat. 5 Ethernet cable. The service port is located at the front of
the CIU 888 (see FIGURE 2-10).
CIU 888: Service port
b) Unlock both the primary CIU 888 and secondary CIU 888 to
enable changes in the configuration using the key switches
located at the front of the CIU 888 (see FIGURE 2-2).
CIU 888: W&M lock key switch (left) and Configuration lock key switch (right)
If the W&M lock key switch has been sealed, the hardware seal (wire) must be removed first in
order to unlock the key switch. By breaking the hardware seal, the CIU 888 will no longer be
W&M approved.
Insert the key for unlocking in the key switch, and turn the key
clockwise.
Remove the key from the key switch and store it.
Firmware Upgrade Manual
CIU 888
2 - 9

Advertisement

Table of Contents
loading

Table of Contents