Step
Operation
9
100 % Calibration
Adjust 100 %
Calibration
Save 100 %
Calibration
10
0 % Calibration
Adjust 0 %
Calibration
Save 0 %
Calibration
11
Return to Online
Display
9/11
Press
AdJH
Momentarily
MODE
AdJH (flashes) – You are now in EDIT
Hold
mode
Range: 19.60 to 20.40 mA typically
(default 20.00 mA)
+AdJH (increments value)
AdJH (decrements value)
Momentarily
MODE
AdJH
Momentarily
AdJL
Momentarily
MODE
AdJL (flashes) – You are now in EDIT
Hold
mode
Range: 3.80 to 4.40 mA typically (default
4.00 mA)
+AdJL (increments value)
AdJL (decrements value)
Momentarily
MODE
AdJL
Momentarily
MODE
Returns to Online Display
Momentarily
DirectLine® DL423 Sensor Module User Manual
Configuration
Display
17