Honeywell DirectLine DL424 User Manual page 23

Sensor module for dl 5000 dissolved oxygen probes
Table of Contents

Advertisement

Step
Operation
5
Enter Output
Configuration
100% Range Value
Selection
Edit 100% Range
Value Selection
Select desired
100% DO Range
Save the New 100%
Range Value
6
0 % Range Value
Selection
Edit 0 % Range
Value Selection
Select 0 % DO
Value
Save the New 0 %
Range Value
7
100 % Calibration
Adjust 100 %
Calibration
Save 100 %
Calibration
9/11
Press
MODE
Momentarily
Momentarily
MODE
Hold
Momentarily
MODE
Momentarily
Momentarily
MODE
Hold
Momentarily
MODE
Momentarily
Momentarily
MODE
Hold
Momentarily
MODE
Momentarily
DirectLine® DL424/425 Sensor Module User Manual
Display
OutC Enter Output Calibration
rnGH (for 1 second) then,
(Value of current 100 % Range Value
Selection)
Flashing Display – You are now in EDIT
mode
(Value of current 100 % selection)
Selected 100% DO Value
Range:
DL424 = 0 to 20 ppm
DL425 = 0 to 200 ppb
(New Value)
rnGL(for 1 second) then,
(value of current 0% Range Value
Selection)
Flashing Display – You are now in EDIT
mode
(value of current 0 % selection)
Selected 0 % DO Value
Range:
DL424 = 0 to 20 ppm
DL425 = 0 to 200 ppb
(New Value)
AdJH
AdJH (flashes) – You are now in EDIT
mode
Range: 19.60 to 20.40 mA typically
(default 20.00 mA)
+AdJH (increments value)
AdJH (decrements value)
AdJH
Configuration
17