Honeywell SPM Flex User Manual page 50

Single point monitor gas detector
Hide thumbs Also See for SPM Flex:
Table of Contents

Advertisement

SPM Flex Gas Detector
Display
Type Sub-
type
String
101
Flow Failure
103
Gate Motor Failure
104
Tape Advance Failure
105
High Background
106
Gas Table Invalid
107
Comm� Failure RFID�
108
Chemcassette Read Failure
109
Temperature Too High
110
Temperature Too Low
111
Optics Error
112
Low Voltage Fault
113
Chemcassette Expired
114
Optics Subsystem Fail
115
Comm� Failure GD
117
NV Memory Corrupt
118
mA Output Mismatch
119
Ethernet Fail
120
Battery Very Low
1
Flow Warning
2
Chemcassette Low
3
Idle Timeout
4
Low Voltage Warning
5
Comm Failure SPI
6
File System Error
7
Comm� Invalid GD
8
Warn CC Expiring
9
Software Error
10
Optics Adjust Fail
11
Flow Cal Fail
12
Inhibit Timeout
13
Force mA Timeout
14
Force Relay Timeout
15
Battery Low
50
Instrument Fault/Maintenance Fault/Information Codes
Technical
Description
Flow failure
Gate timeout
Tape advance timeout
optics counts too low after multiple pulls
Gas table invalid
comm� fail to RFID reader
RFID read failure
overtemperature
undertemperature
value of LED drive or reference detector signal
out of range
DC input voltage low
Now>LifeInBag+Mfg or
Now>LifeOutOfBag+FirstUse
no comm� to optics block
no comm� to GD
CRC error
Feedback ADC mismatches DAC
Ethernet hardware in UI failed to initialize
Battery critically low, shutting down
Flow is unregulated
Tape will run out in 24 hrs
out of monitor too long
DC input voltage low
no SPI comm� to UI
File system error in NAND or USB�
Invalid SPI parameter data from GD�
A stored tape will expire in 7 days; if enabled,
the same fault will be issued for cartridges
in use�
catch all
Calibration failure
Calibration failure
Timed out
Timed out
Timed out
Battery Low
Meaning of
Probable
Parameter
Cause
Pump failure
-Flow line obstruction
Flow, cc/min
-Positive pressure
Gate motor failure
-Gate motor sensor
N/A
failure
Broken tape
-Tape drive motor
failure
N/A
-Encoder failure
N/A
Tape issue
error code
CC RFID issue
N/A
RFID PCB failure
N/A
CC RFID issue
Environment is
Temperature, C
over recommended
temperature
Environment is
Temperature, C
below recommended
temperature
N/A
Optics PCB failure
voltage in Volts
Faulty power circuit
Chemcassette
1�0 or 2�0 for two causes
cartridge has expired
Optics cable
disconnected
N/A
-Optics PCB failure
N/A
Main PCB failure
code to which test failed�
PCB failure
Faulty circuit
Incorrect external
Error magnitude in mA�
wiring
Bad calibration
External connection
issue
N/A
Main PCB failure
Battery level
Low battery
Flow out of control
Flow, cc/min
range
N/A
Low tape
N/A
Out of monitor
voltage in Volts
N/A
N/A
N/A
N/A
Old tape
error code
error code
Calibration failure
error code
Calibration failure
Detector has been left
N/A
in inhibit mode longer
than timeout warning
Detector has been in
N/A
force mode too long
Detector has been in
N/A
force mode too long
Battery level
Battery is low
Corrective
Action
Replace pump
Check flow lines
Check for positive
pressure
Replace
Replace
Check, replace
Replace, contact HA
Replace
Replace, contact HA
Improve ventilation/
cooling
Bring environment to
correct temperature
Replace
Contact HA
Replace
Reconnect, replace
Contact HA
Replace
Contact HA
check wiring
Perform mA
calibration
Check external
Ethernet connection
Replace
Charge detector
Check Chemcassette
Check sample tube
Replace
Put in monitor and
reset fault
Contact HA
Contact HA
Contact HA
Replace
Contact HA
Perform calibration
again, Check the
calibration setup,
Contact HA
Perform calibration
again, Check the
calibration setup,
Contact HA
No action required
No action required
No action required
Charge battery
continued���