SPM Flex Gas Detector
Battery
The detector's battery level is indicated from 0 (fully discharged) to 100%
(fully charged)� The battery icon displays the approximate battery level on the
status bar, while a more accurate value can be found in the Review mode under
Additional Status� If the battery's charge is too low for safe operation, a "Critically
low battery!" message will be displayed and the detector will turn off�
Bar graph
The bar graph displays the concentration reading up to double the Alarm 2
value for the gas being monitored� The Alarm 1 and Alarm 2 values are indicated
numerically on the bar graph�
The display can be configured for different backlight and dimming options� By
default the backlight will dim after a few minutes of no activity� The display can
be configured to turn off the backlight entirely after a set period of time�
Inhibit
When the detector is in inhibit mode, it will show a bell icon with a red slash
through it�
1. Use the [Up] or [Down] buttons to select the "Inhibit" on the Maintenance menu.
2. Press the [Accept] button.
3. The Inhibit Type menu or the Time Out menu can then be selected. Possible
inhibit types are none, alarms only, faults only, alarms and faults, and all, as
shown in the following table). The Time Out options (the time until the detector
exits inhibit mode and returns to active monitoring) are from 1 to 60 minutes.
4. To take the detector out of inhibit mode, select "None" from the Inhibit Type
menu and press the [Accept] button twice to return to the Maintenance menu.
NOTE
If the inhibit times out before the inhibit state is returned to "none,"
maintenance fault code M17 will be displayed�