95 MPD-IF1
Characteristic
Gases Detected: Propane gas in the %LEL range.
Measuring Range: 0 – 100%LEL Propane
Operating Temperature Range: -20°C to +50°C
Operating Humidity Range: 0 to 95% RH non-condensing
Operating Pressure Range: 80 to 110 kPa
Warm Up Time: 3 minutes
Response Time: T90 = <30 seconds
Voltage Range: 3.0V ~ 5.0V (From control equipment)
Drive Current: 75mA ~ 85mA (From control equipment)
Power Consumption: <500mW
Signal Output: mV Bridge
Zero Stability with Time: <±3% FS/year
Span Stability with Time: <±5% FS/year
Span Stability with
Temperature:
* May not be applicable when using gas cross-
Expected Operating Life: 5 years
30
Combustible gases in the %LEL range with sensitivity
dependent on gas type.
Other gas ranges may vary.
Time to reach 90% of a step change in gas concentration of
100% of FS Propane
Linearity: <±5% FS
Accuracy: <±5% FS
<±10% of reading up to 50% FSD, ±15% of reading from
50% to 100% FSD or 2% FSD whichever is greater over the
operating temperature range.
reference factors
Dimensions: 2.080 in Dia x 3.016 in L / 52.83 mm Dia x 76.62 mm L
Weight: 1.2 lbs / 544.3g