Lcd Z-Lock Replacement - GE Vivid E80 Service Manual

Hide thumbs Also See for Vivid E80:
Table of Contents

Advertisement

LCD Z-Lock replacement

Read and follow
WARNING
Preparations
LCD Z-Mount Lock removal procedure
NOTE:
LCD Z-Mount Lock installation procedure
Vivid E80/E90/E95
GC091052 Rev. 3
LCD Monitor and LCD Arm parts replacement
Energy Control and Power Lockout for Vivid E80/E90/E95.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
2. Disconnect the Mains Power Cable from the wall outlet.
3. Disconnect all probes and external I/O cabling.
4. Remove the Bulkhead Cover.
ADD RELATED INFO
It is not necessary to remove the LCD Monitor or the LCD Arm.
Follow these steps to remove the LCD Mount Lock:
1. Remove the two screws securing the LCD Z-Mount Lock.
2. Remove the LCD Z-Mount Lock.
Follow these steps to install the LCD Z-Mount Lock:
1. Position the LCD Mount Lock.
2. Install the two screws securing the LCD Mount Lock.
3. Install the Bulkhead Cover.
Service Manual
8-33