Reassembling the Printer Assembly

2031504-159D
8. Remove the printer motor from the printer mounting base.
1. Replace a new printer motor on the bottom assembly as shown.
2. Replace the two mounting screws.
3. Reconnect the printer cable to the mainboard.
4. Replace the printer door.
5. Reassemble the LCD assembly.
6. Reassemble the top cover assembly.
7. Reassemble the battery assembly.
8. Perform the applicable checkout procedures.
Refer to
"Functional Checkout"
MAC™ 800
on page 4-24.