Nibp Calibration - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

4.7.2 NIBP Calibration

Active Leak Test: Wrap an adult cuff around a pipe
and connect the cuff to the module. Select the
active leak test (ON). The module automatically
pumps a pressure of 260 mmHg into the cuff. Wait
for several seconds until the pressure stabilizes.
Then check that the pressure reading does not drop
more than 6 mmHg per minute. If it does, leaking
point(s) should be detected and fixed. Cancel the
test by selecting the Active leak test OFF.
Calibration Check: After the calibration check is
selected (ON), the module zeroes the pressure
transducers at the beginning of the calibration
check. Do not pump pressure until the text
'Calibrating' appears in the NIBP digit field or the
zeroing will fail. After the zeroing is done, manually
pump pressure into the module and make sure that
the same pressure values are shown both on the
display and on the manometer. Pressure of both
pressure channels B1 and B2 are shown. The
pressure values are automatically zero-
compensated, so the readings of B1 and B2 should be the same as the manometer readings.
Protection: Software calibration protection (ON/OFF). Select
becomes available in the menu after pressing the NIBP module buttons
simultaneously for 3 seconds. For M-NE(12)STPR series modules: Protection can be set ON or OFF only when
the toggle switch at the bottom module is set to the right.
How to Calibrate
Calibration selection is available only when protection is OFF.
NOTE: Both channels B1 and B2 must be calibrated simultaneously.
NOTE: The module must be in the frame during the whole procedure.
NIBP calibration can be performed in the NIBP Service menu as follows:
1.
If Protection is ON, change it to OFF by pressing the NIBP module buttons
and
Then press the buttons again for 3 seconds to enable Calibration.
NOTE: When the buttons have been pressed, the NIBP field shows an error message
'Calibration switch on!'.
NOTE: For M-NE(12)STPR series modules: If Protection is ON, change it to OFF by first turning
the toggle switch to the right at the bottom of the module, which enables the Protection
selection. Then turn the toggle switch to the left to enable Calibration. When the switch is at
the right, the NIBP field shows an error message 'Calibration switch on!'
NOTE: When calibration is enabled, a message 'Calibration not protected' appears.
2.
For proper zeroing to take place, remove the hose from the front panel connector. Select
Calibration and push the ComWheel. Messages 'Zeroing' and 'Zeroed' will appear in the
NIBP message field. After this, a pressure bar will appear beside the menu and the text
'Calibrating' will appear in NIBP digit field.
Start Cancel
simultaneously for 3 seconds, which enables the Protection selection.
OFF
Protection
when calibrating.
Auto ON/OFF
Document no. M1144955-001
Service Menu
selection
Start/Cancel
and
Auto ON/OFF
35