Track Test; Figure 5-5. Typical Spectrum Test Displays - Honeywell Chadwick-Helmuth 8500C Maintenance Manual

Balancer/analyzer system
Table of Contents

Advertisement

Figure 5-5. Typical Spectrum Test Displays.

h.
Use left or right arrow keys to move cursor (dotted line) to 900 RPM.
i.
Verify that readout at top of display shows peak signal is 0.4 +0.04 IPS at 900 +5 RPM.
J.
Switch Calibrator to 1800 RPM CAM RATE.
k.
Press START key.
l.
Verify that spectrum is displayed (examples in Figure 5-5).
m.
Use left or right arrow keys to move cursor (dotted line) to 1800 RPM.
n.
Verify that readout at top of display shows peak signal is 0.8 +0.08 IPS at 1800 +5 RPM.
5.4.1.5

Track Test

a.
Press TRACK key on 8500C/C+.
b.
Press MORE soft key to select 4 blades.
c.
Press START key.
d.
Aim Strobex at rotor disk on Calibrator 11A and pull trigger.
e.
Verify that stem of "h" at center of rotor disk indicates the following pattern:
f.
Release trigger on Strobex.
1800 RPM CAM RATE
STEM OF "h"
FILE: CROSS.DRW
DATE: 12/10/90
8500C/8500C+ System Maintenance Manual
Functional Tests
5-9