α
PT100
= 0.00385
α
PT200
= 0.00385
α
PT500
= 0.00385
α
PT1000
= 0.00385
100 ohm Nickel
120 ohm Nickel
Cu10#
Cu53#
Reference Temperature: 22°C
Reference Humidity:
65% RH +/-15%
* Does not includes reference junction calibration of ±1.0 º C using the standard "ice bath" method of calibration. Factory accuracy can be im-
proved by performing a field calibration. Also does not include any error on the sensor.
** Tolerance for these input types includes that of the external shunt resistors (0.1% tolerance)
*** Reference Accuracy can be improved to +/- 0.4ºC/0.7ºF using the single point compensation calibration.
# not available on the QXe. Including all Field Cal values.
Specification
Pulse Input (optional)
(not available on the QXe)
Alarm Outputs (optional)
Analogue Outputs
(Re-transmission Outputs)
(optional)
(not available for the QXe)
Digital Input/Output
(optional)
Transmitter Power (optional)
Agency Approval (optional)
Vutronik Connection
(optional) QX Recorder only
Extended Security System
(ESS) (optional)
Miscellaneous
264
-328 to 1562
-200 to 850
-328 to 1562
-200 to 850
-328 to 1562
-200 to 850
-328 to 1562
-200 to 850
-76 to 356
-60 to 180
-112 to 500
-80 to 260
-328 to 500
-200 to 260
32 to 302
0 to 150
4 isolated inputs per board, frequency – 1Hz to 25kHz, updated once per sec. Input: Low < 1V, High >4V
to <50V or Volt free input: Low = short circuit, High = open circuit.
Programmable alarm set points (6 per pen) can be configured to activate up to 16 outputs for the QX and
48 outputs for the SX. Update rate: 200 ms for all alarms. Number/Type:
• 4 or 8 relay contacts SPDT, 3A 240VAC, 3A 24VAC/DC, 0.2A 240VDC (non-inductive, internally
suppressed)
• 8 I/O or 16 I/O - SPNO 1A 24VDC (non-inductive, internally suppressed). 16 I/O not available for
the QXe
Activation: Fully programmable internal alarm levels. Assignable to any relay or discrete output.
2 or 4 re-transmission outputs available for the QX Recorder and 2, 4, 6 or 8 re-transmission outputs avail-
able for the SX Recorder; a pen drives each output. Analog inputs, totalised values or any mathematical
result can be re-transmitted.
Update Rate: 250 msec all channels
Type: 0 to 20 / 4 to 20 mA
Resolution: 0.002%
8 I/O or 16 I/O: all channels may be selected freely as either digital inputs or outputs. The Digital I/O card
also has 4 channels that can be set as pulse inputs (channels 1 to 4). The operating frequency for pulse
inputs on the Digital I/O card is 1kHz max. Input: Low < 1V, High >4.5V to <10V DC (9V to 20V p-p AC) or
Volt free input: Low = short circuit, High = open circuit. 16 I/O not available for the QXe
4 relay outputs: all four channels are relay outputs only.
8 relays/ 2 DI card: two outputs can be configured for use as digital inputs: A digital input is provided by a
volt free contact between the normally open (NO) and the common (C) terminals of an output relay. If the
2 Digital inputs are used only 6 relay outputs are available. Closed <500 ohms, Open >300 kohms.
QX has 200mA @ 24VDC ± 3VDC. SX has 1 Amp @ 24VDC ± 3VDC.
QXe is available with 130mA @ 24VDC TXP as an option on the Comms card.
CSA (Optional) CSA22.2-No.1010.1-2004 Certificate Number 211230. UL (Optional) ANSI/UL61010-1-
2004 File # 201698. FM Class 1 Division 2 (optional). FM not available for the QXe
Optional rear cover with 50-pin connector for direct connection to the recorder. TDC2000/3000 system
using Vutronik Trend Recorder, 24VDC instrument power only.
Provides full support for 21 CFR Part 11. Includes features for entry of unique User ID's and associated
passwords, timeout on inactivity (1 to 10 min.), password expiration (1 to 365 days), up to 50 users, pass-
word re-entry lock out for incorrect entry of password more than 3 times, no re-use of passwords (pro-
grammable 4 to 12 times), traceability by user name
Optional customer ID Tagging (3 lines of up to 22 characters each line)
1.1
0.6
1.1
0.6
1.1
0.6
1.1
0.6
0.9
0.5
0.5
0.3
5.5***
3***
0.5
0.3
Reference Sample Rate: 2Hz (500msec)
Long term stability:
0.2%/year
Options
Accuracy: ±0.1% 0-500
Maximum Load Resistance: 1000 Ohms
Isolation: 300VAC
0.01%/ °C
0.01%/ °C
0.01%/ °C
0.01%/ °C
0.01%/ °C
0.01%/ °C
0.01%/ °C
0.01%/ °C
Ω
Ω
load, +/-0.25% 500
1K
43-TV-25-30 Iss.6 GLO Aug 07 UK
Ω
load