INSTALLATION
3
• Attach and tighten the locking screw.
i
End of the procedure.
INFORMATION!
If the antenna diameter is larger than the process connection, refer to Process connections on
page 32
.
INFORMATION!
If it is necessary to send an order for the orientation system, refer to Accessories: process
connection kits on page 177

3.9.2 How to tilt the device

If you must measure a product which does not have a level surface then attach an orientation
system and tilt the device to get a good measurement signal. You can tilt the device to a
maximum angle of 30° from the vertical axis.
Figure 3-18: How to tilt the device
Equipment needed:
• 5 mm Allen wrench (not supplied)
• Loosen the 4 socket head screws around the orientation system with a 5 mm Allen wrench.
• Tilt the device to the correct angle.
• Tighten the 4 socket head screws. This step locks the orientation system to prevent
movement.