Calibration Troubleshooting; Bump Test Troubleshooting - Honeywell GasAlertQuattro Operator's Manual

1, 2, 3, and 4 gas detector
Table of Contents

Advertisement

GasAlertQuattro
Operator's Manual

Calibration Troubleshooting

Error Screen
H
S ppm
2
0.0
20.9
O
%
2

Bump Test Troubleshooting

For bump test troubleshooting recommendations, refer to Table 1, Startup Troubleshooting.
42
Shop for Gas products online at:
Problem
Detector
Calibrate the sensor.
deactivates while
trying to calibrate
Detector does not
enter calibration.
Displays OFF and
OFF
then deactivates.
Self-test Fail
Replace the sensor or
contact BW Technologies
If a sensor(s) fails
by Honeywell. Refer to
auto zero, an error
CO ppm
Replacing the Sensors on
0
message displays
page 26
showing which
sensor(s) failed.
LEL %
CanarySense
www.
Table 2:
Solution
Error Screen
Cal IR Lock Enabled
IR lock enabled
screen displays.
Calibration
IR lock enabled...
Connect to
MicroDock or
Fleet Manager
Press
to end
waiting... 24
Calibration Error
Insufficient amount of
Calibration
gas detected.
gas not
detected
Press
to skip
Problem
Solution
An IR device must be used
to calibrate (IR Link or
MicroDock II). For manual
calibration, refer to
Calibrating Using an IR
Device in the
GasAlertQuattro
Technical Reference
Guide. For automated
calibration, refer to the
MicroDock II User
Manual.
Verify the span gas values
on the cylinder match the
span gas values set for the
detector. Ensure gas is
applied at a flow rate of
250-500 ml/min. Ensure
cylinder is not empty or
expired. Replace
immediately if required.
Replace the regulator if
required.