Honeywell GasAlertQuattro Operator's Manual page 39

1, 2, 3, and 4 gas detector
Table of Contents

Advertisement

GasAlertQuattro
Operator's Manual
The detector enters alarm
immediately when activated.
The activation startup self-test
fails.
Detector Operation
Detector does not display normal
gas reading after startup
sequence.
36
Shop for Gas products online at:
Problem
Possible Cause
Sensor needs to stabilize.
Low battery or critical low battery
alarm.
Hazardous environment.
A new sensor has been inserted
General fault.
Sensor error.
Sensors not stabilized.
Sensor(s) requires calibration.
Target gas is present.
CanarySense
www.
Solution
Used sensor: Wait 60 seconds
New sensor: Wait 5 minutes
Replace alkaline batteries. Refer to Replacing the Alkaline
Batteries on page 22.
Refer to Charging the Rechargeable Battery Pack on page 21.
Leave the area immediately. Deactivate and reactivate the
detector in a safe area that is free of hazardous gas, in an
atmosphere of 20.9% oxygen.
Calibrate the sensor.
Contact BW Technologies by Honeywell.
Refer to Startup Troubleshooting on page 40. If required, refer to
Replacing the Sensors on page 26.
Used sensor: Wait 60 seconds
New sensor: Wait 5 minutes
Refer to Calibration on page 10.
Detector is operating properly. Use caution in suspect areas.
.com
1.888.610.7664