Honeywell GasAlertQuattro Operator's Manual page 21

1, 2, 3, and 4 gas detector
Table of Contents

Advertisement

GasAlertQuattro
Operator's Manual
• Force Calibration: If enabled, during startup if a sensor(s)
is past due for calibration, the sensor(s) must be calibrated
to continue and enter normal operation.
A value must be entered in the Cal Interval (days) field
before enabling Force Calibration.
• Force Bump: If enabled, during startup if the sensor(s) is
past due for a bump test, a bump test should be performed
and the overdue sensor(s) must enter into alarm.
A value must be entered in the Bump Interval (days) field
before enabling Force Bump.
• Cal IR Lock: If enabled, the sensor(s) can only be
calibrated using an IR device (IR Link or the MicroDock II
station).
If the Cal IR Lock option is enabled and a manual
calibration is attempted, the sensor(s) will auto zero
but they will not be calibrated.
Do not use for gas detection while connected to a PC.
• Flip Display: The detector can display screens at 0°
(upright) or 180° (upside down), depending upon how the
detector is worn by the worker. If the Flip Display option is
enabled, the LCD is viewed at 180° (upside down).
18
Shop for Gas products online at:
Note
CanarySense
www.
• Stealth: When enabled, the following features are
disabled: backlight, audible alarms, visual alarms,
IntelliFlash, and confidence/compliance beep. Only the
vibrator and the LCD readings activate during an alarm
condition.
Note
Should the user wish to comply with European
performance certifications, stealth mode must be
disabled.
• Datalog Interval: The Datalog Interval (seconds) field
defines how often the detector records a datalog (every
1-120 seconds). Enter the desired value.
The total number of 8-hour days datalogs that can be
recorded is assuming 90% of the day has no gas
concentrations.
When the memory is full, the detector replaces the oldest
datalogs with the most recent datalogs.
• IntelliFlash Interval: The IntelliFlash Interval (seconds)
field defines how often (every 1-120 seconds) the
IntelliFlash occurs.
• Confidence/Compliance Beep Interval: Define how
often (every 1-120 seconds) the confidence/compliance
beep occurs.
• Language: The Language field provides a drop down
menu that includes the following language options:
English, Français (French), Deutsch (German), Español
(Spanish), Português (Portuguese). Select the language
from the drop-down menu in FleetManager II.