iDRX-TC SPECIFICATIONS
Type J,K,T,E,R,S,B,N, J DIN
INPUT TYPE:
Input Type
Range °F
J
-346 to 1400°F
K
-454 to 2500°F
T
-454 to 752 °F
E
-454 to 1832 °F
R S
-58 to 3214 °F
B
+212 to 3300 °F
N
-454 to 2372°F
J DIN
-328 to 1652 °F
Dielectric strength per 1 minute test based on
ISOLATION:
EN 61010 for 50 Vdc or Vrms working voltage.
Three way Isolation:
• Power to input 1800V Peak
• Power to analog output/communication 1800V Peak
• Input to analog output/communication 1400V Peak
INPUT OVER-VOLTAGE PROTECTION:
250 V AC for 1 minute
Dual slope
ANALOG TO DIGITAL TECHNIQUE:
3 readings/second, automatic polarity
READ RATE:
±1°C for temperature above -150 °C
ACCURACY AT 25 °C:
±2°C for temperature below -150 °C
0.05 °C/°C
TEMPERATURE STABILITY:
2 seconds to 99% of the final value
STEP RESPONSE FOR RS485 OUTPUT:
(Filter time constant = 64)
To verify the response time, check the carriage return
RESPONSE TIME:
<CR>, it will be sent at the end of the response. You can
send another command after you receive the <CR>.
i.e.
send: *01X01
response: 01X01<DATA><CR>
Note:
01 is address.
30 minutes
WARM UP TO RATED ACCURACY:
10 to 32 Vdc
INPUT POWER:
2.4 Watts (100 mA at 24 Vdc)
POWER CONSUMPTION:
Range °C
-210 to 760 °C
-270 to 1372 °C
-270 to 400 °C
-270 to 1000 °C
-50 to 1768 °C
+100 to 1820 °C
-270 to 1300 °C
-200 to 900 °C
Page 29
-5 to +55 °C
OPERATING AMBIENT:
-40 to +85 °C
STORAGE TEMPERATURE:
90% at 40 °C non-condensing
RELATIVE HUMIDITY:
J 4
J 3
J 2
J 1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Idrx-acvIdrx-fpIdrx-tcIdrx-stIdrx-prIdrx-rtd

Table of Contents