9.4
Electrolyte Concentration Calculation
An example of Na calculation with use of the printout example on the previous page is
described here.
Calculation of SL (slope value)
−
E
E
H
L
SL =
C
H
log
C
L
−
−
32 0
.
( .
39 6)
=
160
log
120
= 60.8
SL :
Slope value
E
:
Electromotive force of HIGH solution (32.0)
H
E
:
Electromotive force of LOW solution (-39.6)
L
C
:
Concentration of HIGH solution (160.0)
H
C
:
Concentration of LOW solution (120.0)
L
Concentration calculation of internal standard solution
−
E I S E L
×
C
= CL
10
SL
IS
−
35 4
×
= 120 0 10
.
= 140.7
C
:
Concentration of internal standard solution
IS
E
:
Electromotive force of internal standard solution (-35.4)
IS
Calculation of compensation value (C.VALUE)
The difference between the input value and measured value is obtained through measurement
of the Hitachi calibrator having the known concentration.
C.VALUE = C
- C
c
x
= 139.0 - 138.9
= 0.1
C
:
Input value (139.0)
c
C
:
Measured value (138.9)
x
− −
.
(
39 6
. )
60 8
.
9 - 13