34
en | Programming
4.4
Special settings
After pressing
4.4.1
Displaying firmware version
1.
2.
V2.01 BN111.240.00A
4.4.2
Resetting all the parameters
CAUTION!
Disconnect the RS-485 bus before performing a reset. When you are finished, the RS-485 bus
can be connected again.
The following procedure resets all the programmed parameters of the NurseCall Main Unit to
the original factory settings. See Section 4.2.1 List of original factory settings, page 23.
NOTICE!
This reset is mandatory if you wish to change the locating mode or the display mode.
See Section 4.2.3 Locating mode, page 24 and Section 4.2.4 Display mode, page 24.
It is also mandatory before setting the NurseCall Main Unit as Universal NurseCall or as
Standard NurseCall. See Section 4.4.7 Standard NurseCall selection, page 35 and
Section 4.4.8 Universal NurseCall selection, page 36.
1.
2.
3.
4.
5.
F.01U.252.699 | V1.1 | 2012.01
three times quickly, you can enter special codes.
Enter Code
Type the code 194155.
The version of the firmware will be displayed for a few seconds.
Software REV A
Type the code 194156.
RESET TOTAL ?
(OK) = YES
A confirmation is required.
Press
to confirm the reset or
When the unit resets, it emits a short melody and displays a temporary message.
Please Wait...
After a few seconds, the unit goes back to the first use display.
Select the language, the locating mode and the display mode.
See Section 4.2 First use.
if you wish to cancel the reset.
User Manual
NurseCall Main Unit
Bosch Security Systems