Service and maintenance
7.3 Maintenance and repair work
Enclosure open
Risk of explosion in hazardous areas as a result of hot components and/or charged capacitors
inside the device.
To open the device in a hazardous area:
1. Isolate the device from power.
2. Observe the wait time specified in Technical specifications (Page 77) or on the warning
3. Visually inspect sensor inlet and outlet.
Exception: Devices exclusively having Intrinsic safety (Ex i) may be opened in an energized
state in hazardous areas.
Improper connection after maintenance
Risk of explosion in areas subject to explosion hazard.
• Connect the device correctly after maintenance.
• Close the device after maintenance work.
Refer to Connecting (Page 61).
Hot, toxic, or corrosive process media
Risk of injury during maintenance work.
When working on the process connection, hot, toxic, or corrosive process media could be
released.
• As long as the device is under pressure, do not loosen process connections and do not
• Before opening or removing the device ensure that process media cannot be released.
Hot surfaces
Risk of burns during maintenance work on parts having surface temperatures exceeding 70 °C
(158 °F).
• Take corresponding protective measures, for example, by wearing protective gloves.
• After carrying out maintenance, remount touch protection measures.