Getting started Introduction Safety notes SITRANS F Description Ultrasonic flowmeters SITRANS FSS100 Installing/mounting Connecting Installation Manual Service and maintenance Technical specifications Dimension drawings Product documentation and support Appendix 7ME381 (SITRANS FSS100) 12/2023 A5E52595000-AA...
Page 2
Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
You have read the following safety instructions: • General safety information (Page 13) • Basic safety information: Installing/mounting (Page 27) • Basic safety information: Connecting (Page 61) Read the entire document in order to achieve the optimum performance of the device. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 8
Getting started 1.1 Getting started SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Introduction FSS100 manual usage Note This manual applies to the SITRANS FSS100 ultrasonic retrofit sensors only. Designated use Use the device in accordance with the information on the nameplate and in the Technical specifications (Page 77). See also Nameplate layout (Page 65)
Checking the consignment 1. Check the packaging and the delivered items for visible damages. 2. Report any claims for damages immediately to the shipping company. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Siemens’ products and solutions undergo continuous development to make them more secure. Siemens strongly recommends that product updates are applied as soon as they are available and that the latest product versions are used. Use of product versions that are no longer supported, and failure to apply the latest updates may increase customer’s exposure...
The content reflects the technical status at the time of publishing. Siemens reserves the right to make technical changes in the course of further development.
Hot surface Dangerous electrical voltage Corrosive materials Toxic materials Isolate the device from power using a circuit-breaker Protect the device from impact otherwise loss of degree of protection Protective insulation; device in protection class II SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Directive of the European Parliament and the Council on the re‐ stances directive RoHS striction of the use of certain hazardous substances in electrical 2011/65/EU and electronic equipment The applicable directives can be found in the EU Declaration of Conformity of the specific device. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Note Operation under special ambient conditions We highly recommend that you contact your Siemens representative or our application department before you operate the device under special ambient conditions as can be encountered in nuclear power plants or when the device is used for research and development purposes.
Page 16
There is a risk of explosion. • Connect the device with type of protection "Intrinsic safety" solely to an intrinsically safe circuit. • Observe the specifications for the electrical data on the certificate and/or in Technical specifications (Page 77). SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
• Mass flow • Total volume or mass SITRANS FSS100 systems are maintenance free, self cleaning and have a service life of more than 10 years. As with the SITRANS FS230 and FS220 clamp-on systems, the SITRANS FSS100 sensors can also be operated with the SITRANS FST020 and FST030 transmitters.
Page 18
The SITRANS FS120 / SITRANS FS130 retrofit sensor kit is used for installing one to four sets of ultrasonic transducers type FSS100 in existing pipes made of steel or concrete. The images below show the position or transducers relative to flow direction. SITRANS FSS100 1-track flow direction SITRANS FSS100 2-track flow direction SITRANS FSS100...
Page 19
Description 4.2 Design SITRANS FSS100 3-track flow direction SITRANS FSS100 3-track alternative flow direction SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
DSL. You can also connect the transmitter to other flow ultrasonic sensors. Wall mount transmitter overview with FS DSL11 Transmitter bottom view External DSL side view A External DSL side view B SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
The SITRANS FSS100 is ideal for installation in a new system. Compared to an ultrasonic clamp-on flowmeter, the sensors are installed directly into the existing pipe system. After installing the sensors and transmitter, you have an uncalibrated inline flowmeter. It is a much cheaper solution than installing an inline flowmeter with a tube into the system.
The proportional flow factor K is determined by wet calibration or calculated by "AUTO" in case of manual programming of mechanical/geometrical pipe data (Retrofit only). The transducer angle (θ), distance between sensors (L) and pipe dimension (D and D ) are shown in the figure below. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
For further details see Ultrasonic certs on SIOS (https:// support.industry.siemens.com/cs/ww/en/ps/17368/cert). The device is available with approvals for general purpose and for hazardous areas. In all cases, check the nameplate on your device, and confirm the approval rating. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Hot, toxic and corrosive media could be released if the wetted parts are unsuitable for the process medium. • Ensure that the material of the device parts wetted by the process medium is suitable for the medium. Refer to the information in Technical specifications (Page 77). SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 28
Note Material compatibility Siemens can provide you with support concerning selection of parts wetted by process media. However, you are responsible for the selection of parts. Siemens accepts no liability for faults or failures resulting from incompatible materials. WARNING Unsuitable connecting parts Risk of injury or poisoning.
• Observe the maximum permissible ambient temperature. Refer to the information in the section Technical specifications (Page 77). NOTICE Aggressive atmospheres Damage to device through penetration of aggressive vapors. • Ensure that the device is suitable for the application. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
• Before installing ensure there is no visible damage to the device. • Make sure that process connectors are clean, and suitable gaskets and glands are used. • Mount the device using suitable tools. Refer to the information in Technical specifications (Page 77). SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Note Material compatibility Siemens can provide you with support concerning selection of sensors. However, you are responisible for the selection of components. Siemens accepts no liability for faults or failures resulting from incompatible materials. Basic installation requirements Installation location The FSS100 sensor can be installed both indoors and outdoors, even in exposed surroundings.
Page 32
NOTICE In case of large temperature differences between the medium and the environment, the transducer must be isolated to avoid 2-phase flow which will result in inaccurate measuring results. Space requirements (empty pipe installation) SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Figure 5-1 Sensor filled with liquid 1-track Figure 5-2 Sensor filled with liquid 2-track Avoid the following installations: • Installation at the highest point of the pipe system • Installation in vertical pipes with free outlet SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Sensor orientation Horizontal (recommended) Note Liquids containing abrasive or other particles: Mount sensor in vertical or inclined pipe position to minimize wear. Vertical The vertical transducers can be oriented in vertical position with upwards flow. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Note Select a valve not altering flow when fully opened. Recommended inlet Fully open valve, ie no flow restriction Min. 10 × pipe diameter Partially open valve Min. 40 × pipe diameter SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
• Angle iron of a length approximately corresponding to the outer diameter of the pipe • Measuring tape • Scriber, pencil or the like • Pocket calculator • Calipers • Sliding gauge • Hammer • Centre punch • Spirit level • Roll of adhesive tape SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 37
The unit is fastened to the pipe wall by means of vacuum or clamping band. • Filler Note A suitable filler or mortar does not shrink when setting. A comparatively short setting time is preferred. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
The measuring tape must fit snugly around the pipe. Read the circumference (C). Repeat the measurement several times to avoid inaccuracies. The wall thickness is obtained by measuring a similar pipe whose wall thickness is accessible, or by consulting the relevant standard. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 39
3. Mark 4 -5 points on pipe top line using this method. 4. Draw a line X - X through the marked top points using length of angle iron. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Up to DN 1000 the transducer position is calculated using the roll of paper supplied with the mounting kit. Figure 5-5 Drawing paper (supplied with the FSS100 up to DN 1000). Note For installations larger than DN 1000, mark directly on the pipe. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 41
1. Draw a line x-x through the middle of the paper. Ensure that the line x-x is right-angled to edge of paper. 2. Mark a point X1 on line x-x approx 5 cm from edge of paper. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 42
3. Draw a line X1 at right angles to x-x. 4. Mark a point A on line X1. Calculate distance a from X1 to A as follows: a = 0.25 x C C = Pipe circumference SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 43
Calculate d using the formula: d = @d x C (for @d see "Correction factors" in the Appendix) Note It is useful to draw a circle around P1 and P2 (∅ = 60 mm, r = 30 mm) SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 44
Calculate b using the formula: b = @b x C (for @b see "Correction factors" in the Appendix) 7. Mark points P3 and P4 by repeating the proceeding procedure The drawing on the paper is now finished. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 45
Fix paper to pipe with adhesive tape 2. Punch points P1, P2, P3 and P4 onto the pipe with centre punch. Ensure that lines dissecting the points can be redrawn once the paper is removed. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 46
When preparing a steel pipe, it is useful to draw a circle with a radius of 30 mm around points P using the centre punch. This facilitates cutting the holes with a flame cutter. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
After marking the top line (see above under 5.3.2) following steps should be done: 1. Mark a point X on top line X - X. 2. Draw a line at right angles to the top line X - X through point X SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 48
C = Circumference of the pipe 4. Mark points P1 and P2 at a distance d from point A. Calculate d using the formula: d = @d x C (for @d see "Correction factors" in the Appendix) SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Step 2: Installing the transducer holders The second step in the installation procedure is to install the transducer holders. In the following it is described how to: • Install the holders on steel pipes SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Or alternatively how to • Install the holders on concrete pipes. 5.10.2.1 Installation on steel pipes Drilling 1. Cut or drill 4 holes centred on P1 and P2. Dimension of the holes: 60 mm +5/-0. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 51
1. Place a set of transducer holders and mounting plates on the mandrel (supplied with the kit) without permanently fixing them. 2. Tack-weld the mounting plates in at least 3 places with transducer holders and mandrels mounted. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 52
Installing/mounting 5.10 Installation instructions 3. Remove transducer holders and mandrel, and weld the mounting plates completely onto the pipe. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 54
3. Tack-weld the correctly positioned transducer holders in at least 3 places with mandrels inserted. 4. Remove mandrel and fully weld transducer holders. Note Make sure that the transducer holders are bent as little as possible by the heat generated during welding. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Making the holes Drill holes at the marked positions, ø 46 mm, into the pipe wall at an angle of 60°. Note It is recommended to use special drill units, see above: Necessary tools (Page 36) SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 56
3. Fix the transducer holder to the pipe by means of mortar. Note Make sure that mandrel is not exposed to mortar as this might result in mandrel getting stuck to transducer holders. Mandrel keeps transducer holders in place until mortar has set. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
④ 2. Connect the wire without a spade terminal to ⑤ 3. Connect the wire with a spade terminal to (the transducer housing). ① 4. Connect the wire of the transducer coax cable to SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 58
• Support the terminal box while loosening the union. Then remove the terminal box from the transducer holder. CAUTION Unsuitable turning of terminal box Risk of device malfunction due to damaged wires inside the terminal box • Do not turn the terminal box SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
• If the device contains hazardous media, it must be emptied prior to disassembly. Make sure that no environmentally hazardous media are released. • Secure the remaining connections so that no damage can result if the process is started unintentionally. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Risk of explosion in hazardous areas as result of open cable inlet or incorrect conduit system. • In the case of a conduit system, mount a spark barrier at a defined distance from the device input. Observe national regulations and the requirements stated in the relevant approvals. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 62
• The cable cross-section of the equipotential bonding cable must be greater than or equal to the connecting cable of the electronics WARNING Unprotected cable ends Risk of explosion through unprotected cable ends in hazardous areas. • Protect unused cable ends in accordance with IEC/EN 60079-14. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Connecting or disconnecting a powered device in explosive environments can lead to an explosion. • Connect and disconnect in non-explosive environments. - or- • Remove power to the device before connecting or disconnecting in explosive atmosphere. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 64
• Connect the device according to the information in Technical specifications (Page 77). NOTICE Ambient temperature too high Damage to cable sheath. • At an ambient temperature ≥ 60 °C (140 °F), use heat-resistant cables suitable for an ambient temperature at least 20 °C (36 °F) higher. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Each FSS100 Ultrasonic flowmeter sensor has a nameplate showing the following information: • product identification • product specifications • certificates and approvals Note Identification Identify your device by comparing your ordering data with the information on the product and specification nameplates. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 66
Connecting 6.3 Nameplate layout SITRANS FSS100 sensor: Nameplate with specific information ① SITRANS FSS100 Product name ② 2813 Notified Body ID (ATEX, example) ③ QR code Product-specific QR code ④ Ex approval specifications ⑤ Manufacturer Manufacturer name and location ⑥...
Page 67
With the use of a smart phone, the QR code provides a direct link to • the product support portal, which includes access to the "How to Install" YouTube video • the product and production-specific documentation maintained in the production database. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
• Ensure that the atmosphere is explosion-free (hot work permit). CAUTION Releasing button lock Improper modification of parameters could influence process safety. • Make sure that only authorized personnel may cancel the button locking of devices for safety-related applications. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Impermissible repair of the device • Repair must be carried out by Siemens authorized personnel only. WARNING Impermissible repair and maintenance of the device • Repair and maintenance must be carried out by Siemens authorized personnel only. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 71
WARNING Impermissible repair of explosion protected devices Risk of explosion in hazardous areas • Repair must be carried out by Siemens authorized personnel only. WARNING Maintenance during continued operation in a hazardous area There is a risk of explosion when carrying out repairs and maintenance on the device in a hazardous area.
Page 72
Risk of burns during maintenance work on parts having surface temperatures exceeding 70 °C (158 °F). • Take corresponding protective measures, for example, by wearing protective gloves. • After carrying out maintenance, remount touch protection measures. SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
• If maintenance measures in an energized state are necessary, observe the particular precautionary measures. Have maintenance work carried out by qualified personnel. Ordering of spare parts Condition • You have a Siemens Industry Mall account. Procedure 1. Open the PIA Life Cycle Portal (https://www.pia-portal.automation.siemens.com). 2. Select the desired language.
5. Select a spare part and add it to your watch list. The watch list opens. 6. Click "Add to cart of Industry Mall". The Siemens Industry Mall opens and you can order your spare part. Replacing the device CAUTION Corrosive substances Risk of chemical burns when replacing the sensor.
Service and maintenance 7.7 Disposal Contact your Siemens representative to clarify if a product is repairable, and how to return it. They can also help with quick repair processing, a repair cost estimate, or a repair report/ cause of failure report.
Page 76
Service and maintenance 7.7 Disposal SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Technical specifications Parts in contact with medium SITRANS FSS100 Material Transducer (Stainless steel) O-ring (FFKM) Transducer holder (Stainless or carbon steel) Max pressure 40 bars Altitude 2000 m above sea level Relative Humidity 95% without condensation Weight SITRANS FSS100 No. of tracks...
Page 78
Technical specifications 8.3 Sizing table m /h l/min. (US) (US) 0.05 0.02 0.01 0.005 10 m/s 1 m/s 0.25 SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Technical specifications 8.5 FS130 wall housing and internal DSL FS130 wall housing and internal DSL SITRANS FSS100 (1 or 2 path-measuring) with SITRANS FST030 wall housing (internal DSL) The SITRANS FST030 transmitter in combination with the SITRANS FSS100 is called SITRANS FS130.
Ex approval SITRANS FST030 Ex Zone 2 (Optional) FS130 wall housing and external DSL SITRANS FSS100 (1 up to 4 path-measuring) with SITRANS FST030 wall housing and FS DSL The SITRANS FST030 transmitter in combination with the SITRANS FSS100 is called SITRANS FS130.
Page 82
0,5 m/s < v < 10 m/s • Without calibration Repeatability ISO 11631 < 0,25% Wall thickness steel pipe Up to 20 mm (0,79") Wall thickness of concrete pipe Up to 200 mm (7,9") > DN 600 SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Ex approval SITRANS FST030 Ex Zone 2 (Optional) FS130 industrial housing and external DSL SITRANS FSS100 (1 up to 4 path-measuring) with SITRANS FST030 industrial housing and FS DSL The SITRANS FST030 transmitter in combination with the SITRANS FSS100 is called SITRANS FS130.
Page 84
Up to 20 mm (0,79") Wall thickness of concrete pipe Up to 200 mm (7,9") > DN 600 Cable length sensor to FS DSL • Min. 3 m • Max. 15 m Connection sensor to FS DSL Analog SITRANS FSS100 Installation Manual, 12/2023, A5E52595000-AA...
Page 85
Min. 5 m • Max. 150 m Connection FS DSL to FST030 Digital (SSL – Siemens Sensor Link) Ex approval SITRANS FSS100 Ex-i oder Ex-d Zone 1 (Optional) Ex approval SITRANS FS DSL Ex Zone 1 (Optional) Ex approval SITRANS FST030...
– "Download as html5, only PC": Open or save the manual in the HTML5 view on your PC You can also find manuals with the Mobile app at Industry Online Support (https:// support.industry.siemens.com/cs/ww/en/sc/2067). Download the app to your mobile device and scan the device QR code. Product documentation by serial number Using the PIA Life Cycle Portal, you can access the serial number-specific product information including technical specifications, spare parts, calibration data, or factory certificates.
(http:// www.siemens.com/automation/csi/service). Service & support on the Internet In addition to our technical support, Siemens offers comprehensive online services at service & support (http://www.siemens.com/automation/serviceandsupport). Contact If you have further questions about the device, contact your local Siemens representative, by doing the following: 1.