Mode! 5344S
Operation and Programming
Description:
The 5344S wiii lock the sweeper to the specified Start Frequency, and then sweep from the specified Start
Frequency to the specified Stop Frequency using the specified Sweep Time. The difference between the
Start and Stop frequency must be less than or equal to 40 MHz. The Sweep time range is 10 ms-100 s.
Connections:
HP-IB
Example:
Start Frequency (STRT) of 2 GHz and a Stop Frequency Stop (STOP) of 2.040 GHz with a Sweep Time of 100
mS.
1.
Set the 5344A HP-IB address to 00 by setting the bottom five switches to the right.
2.
Set the 5344A to the System Controller mode by setting the top HP-IB switch to the left (SYS CONT.)
3.
Set the 8350A address to 00 (equal to the 5344A) by pressing the following key sequence:
SHIFT
(blue key)
j LCL
f
. N
0
f
*\
GHz
$
«,
>
4.
Set MANUAL LOCK and AUTO LOCK off (lamp off) and verify the 5344A front panel CONT lamp is
on. Also verify the plug-in's CW Filter is off.
5.
Using the 5344A Function keys, press and release MODE until the START/STOP annunciator lights.
The FI key will light indicating the 5344A is waiting for a START frequency value. The MODE key will
scroll through the four modes of operation. If you pass STRT/STOP, continue pressing MODE until
you return to STRT/STOP. The FI key will light.
3-44
Figure 3-14. Start/Stop in AUTO LOCK