Model 5344S
Operation and Programming
TRIGGER SWEEP
TS
CD
Initiates a Single Sweep. The TR1 message must have been previously sent, or must proceed the
TS command in the command string.
If the 5344S is in CW mode, sending the TS command is essentially the same as sending the Reset
(RS) command. The 5344S will restart the locking process.
Group Execute Trigger (GET), such as the TRIGGER command (HP 85-Basic) may also be used to
initiate a sweep. Refer to the specific controller's operating manual for more details.
SERVICE REQUEST MASK
RM
Set the 5344A Service Request mask. The RM command determines which of the following
conditions will generate a Service Request.
The RM command is followed by a single digit which specifies the set of conditions that is to
generate a service request:
0-NONE
1 -OUT OF LOCK
2 -END OF SWEEP
3 -OUT OF LOCK or END OF SWEEP
Only one condition may be active at any one time. If more than one is specified, such as RM12,
the second value will be ignored. Notice however, that RM3 considers two conditions to
generate an SRQ interrupt.
3-162.
Input Code Format
3-163.
The 5344A will accept command strings in either upper or lower case. Spaces, carriage
returns, linefeeds and invalid message sequences are ignored. For example, (in HP 85A-Basic):
OUTPUT 701; "CW2GZLK1" or output 701; "cw2gzlk1"