1.2.13
1.2.14,
Environmental Capabilities
E 1-12
The following environmental data are valid only if the instrument is checked in accordance with the
official checking procedure. Details on these procedures and failure criteria are supplied on request by
the PHILIPS organization in your country or by PHILIPS
International
B. V., Industrial & Electro-
acoustic Systems Division, EINDHOVEN, THE
NETHERLANDS.
ambient temperature
—
reference value
—
nominal working range
—
limits for storage and transport
relative humidity
—
reference range
—
nominal working range
—
limit range of use
—
limit range for storage and transport
air pressure
—
reference value
—
nominal working range
air speed
—
reference value
—
nominal working range
heat radiation
vibration
—
limits for storage and transport
operating position
warm-up time
Safety- and Quality Data; Cabinet
protection type
(see DIN 40 050)
protection class
(see IEC 348)
line connection
call rate
radio interference voltage
radio interference radiation
overall dimensions
width
height
depth
weight
+239C +1K
+
59°C...+40°C
-—209C...+70°C
45%...75%
20%...80%
10%...90%
0%...90%
1013 hPa (= 760 mm Hg)
800 hPa... 1060 hPa (= 600... 800 mm Hg; up to 2200 m
height)
Om/s...0.2 m/s
Om/s...0.5 m/s
direct sunlight radiation not allowed
max. amplitude 0.35 mm (10 to 150 Hz) max. 5g
normally upright on feet (horizontal position) or
with bow fold down
30 min.
IP 20
class |, protective conductor
mains connector
< 0.15
acc. to VDE 0871
class B ("'Grenzwertklasse B'')
19", 2 E high
440 mm
105 mm
430 mm
10 kg