Setting And Calibration - Toshiba LF620 Instruction Manual

Electromagnetic flowmeter converter
Hide thumbs Also See for LF620:
Table of Contents

Advertisement

○ Operation timeout function
If no operation is made for one minute or more while the converter is in the setting mode, the
mode automatically returns to the measurement mode unless the parameters are displayed on
the screen.
Menu screen
Parameter check screen
Parameter change screen
(Parameters
changed.)

7.3.2 Setting and Calibration

In the setting mode, you can select items, or check or change the setting values as described below.
When you push
to move the digit, you can execute continuous operation by holding down the relevant
switch longer. (Holding down the switch longer automatically executes the operation continuously.)
● Moving to the menu screen
Switch operation
SET 
A:DETECTOR 
1 EXC CUR 
2 SIZE 
3 EXC FREQ 
4 FLOW DIR 
5 EXIT 
EXC CUR 
 
 
         A 
 
ESC
EXC CUR 
 
 
are
being
         A 
or
to scroll up or down the numeric value or alphabet, or when push
Display example
PUSH SW 
CNT: 
 CNT CTRL 
SET: 
 
 SET MODE 
 
CNT
ESC 
 
MENU SEL 
 
BASIC 
DETAILED 
PREVIEW 
ZERO ADJ 
PLS SET 
ESC
 
When no operation is made
for 1 minute, the mode
automatically returns to the
measurement mode.
ESC
0.2000 
The screen does not return
to the measurement mode.
0.2000 
Mode change screen
SET
Pushing
SET
takes you to the menu configuration
selection screen.
For
configuration,
DETAILED menu.
 
− 45 −
6 F 8 A 0 9 2 7  
Description
in the measurement mode
select
BASIC
or