Toshiba Aquilion TSX-101A/I Series Site Planning Manual page 10

Table of Contents

Advertisement

(7)
Site environment
(a)
Ambient temperature, humidity, and calorific value
Table 4-1 Heat generation by units and environmental conditions
Names of rooms
and units
1. Scanning room
• Gantry
• Patient couch
2. Scan control room
• Console
*1 Maximum heat generation is the heat that is generated when continuous scanning is
performed at the maximum output rating of the unit.
*2 Normal heat generation is the heat that is generated when scanning is not
performed.
*3 1 kW = 860 kcal/h, 1 cal = 4.19 J
*4 A temperature from 0°C to 40°C is permissible if the units are not operated.
(b)
Vibration
0.98 m/s
(c)
Floor construction
Reinforced concrete structure
(d)
Explosion-proof structure
This system is not explosion-proof. Therefore, do not use flammable or explosive gases
near the system.
(e)
Other
Check for evidence of excessive dust or dirt.
Maximum heat
Normal heat
generation *1
generation *2
[kW] [kJ/h] *3
(10.5)
(37,800)
10.0
36,000
0.5
1,800
(8.0)
(28,827)
8.0
28,827
2
(0.1 G) or less
[kW] [kJ/h]
Temperature *4 (°C)
(4.3)
(15,480)
18°C to 28°C where
median value is 20°C
4.0
14,400
to 26°C. Fluctuation
0.3
1,080
around the median
not to exceed ±2°C.
(8.0)
(21,600)
16°C to 28°C
8.0
21,600
10
No. 2A201-519EN*A
Environmental conditions
Relative humidity (%)
40% to 80%
(No condensation)
40% to 80%
(No condensation)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aquilion tsx-101a/h series

Table of Contents