8.4

Electrical Equipment

Electrical work should be executed according to the procedures below:
(1)
Mount an "X-ray ON" lamp (100 VAC or 200 VAC) indicating "X-ray exposure in progress"
above the entrance to the scanning room and connect the power cable of the lamp to the
corresponding power supply facility (100 VAC or 200 VAC) of the site. At this time, connect a
250-V/0.5-A fuse in series to the power input line. In addition, route the cable as described
below to turn ON the lamp in synchronization with the desired CT system operation.
Connection must be performed during system installation.
<1> To turn ON the lamp in synchro-
nization with the power ON operation : Route the cable near the console.
<2> To turn ON the lamp in synchro-
nization with the X-ray exposure
start operation
Lamp type: Incandescent
(2)
100-VAC outlet for maintenance and inspection
Provide two to four 100-VAC outlets each in the scanning room, operating room, and
machine room for use during maintenance and to supply power to facility tools.
8.5

Illumination

Freely adjustable or two-step-adjustable illumination should be installed in the scanning room.
Place illuminating fixtures so that light does not reflect on the display monitor of the console.
Illumination should be adjustable in the range from 10 to 300 lux. In addition, the machine room
must have illumination required for maintenance.
8.6

Cable Connections Between the Units

Figures and tables detailing the cable connections, the inter-unit cables, and the distances
between units are provided on the subsequent pages.
Also refer to subsection 7.5.4 "Pit layout example" when planning the layout.
: Route the cable near the gantry (use the X-ray
system duct for cabling).