Table of Contents

Advertisement

3-2-3. Cell Trouble Without System Error

Check the connection of the harness
between the cell and signal board
and then connect it correctly or replace
it.
Check points:
Flexible Flat Cable (FFC) terminal
surface is not broken.
FFC is connected correctly.
(not skew insertion)
Connector pin is not bent.
Is the symptom improved?
Yes
End
ZRD-1/ZRD-2
,
( 1)
Flexible Flat Cable
terminal surface
Replace the cell.
No
(The reading/writing of the data is
also necessary.
Is the symptom improved?
( 2)
)
No
Yes
End
( 1):
UC-1 board or UP-13 board
( 2):
Refer to the Manual for Color Uniformity
Alignment Software, Section 3-2.
( 3):
Refer to the Manual for Color Uniformity
Alignment Software, Section 3-3.
Replace the UC-1 board.
(The reading/writing of the data is
also necessary.
)
( 3)
End