Wall-Mounted Device; Device-Specific Safety Instructions - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

4.5

Wall-mounted device

4.5.1

Device-specific safety instructions

Open cable inlet or incorrect cable glands
Danger of explosion in hazardous areas
• Close the cable inlets for the electrical connections.
NOTICE
Air circulation around the device
Device damage
The device may overheat in case of heat build-up
• Ensure that there is a sufficient circulation of air around the device.
• If you are installing more than one wall-mounted device, there should be a clearance of
Devices in explosion-proof design
Compact Operating Instructions, 05/2018, A5E35134047-03
WARNING
Only use cable glands or plugs which are approved for the associated type of protection.
≥ 5 cm between the sides of the devices.
Installing
4.5 Wall-mounted device
27