Installing; General Safety Instructions - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Installing

4.1

General safety instructions

Parts in contact with sample gases unsuitable for sample gas
Danger of injury or poisoning, device damage
Hot, poisonous and corrosive sample gases can be released at the connections if the
device parts and accessories coming into contact with sample gas are unsuitable for the
sample gas.
• Use only such connecting part (pipes, threaded joints and sealing material) coming into
Note
Material compatibility
Siemens can provide you with support concerning selection of analyzer components wetted
by the process medium. However, the device operator is responsible for the selection of
components. Siemens accepts no liability for faults or failures resulting from incompatible
materials.
Violation of the maximum permissible operating pressure
Danger of injury or poisoning.
The maximum permissible operating pressure depends on the device version. The device
can be damaged if the maximum permissible operating pressure is exceeded. Hot, toxic
and corrosive process media can be released.
• Make sure that the device is suitable for the maximum permissible operating pressure of
Devices in explosion-proof design
Compact Operating Instructions, 05/2018, A5E35134047-03
WARNING
contact with sample gas that are suitable for the connection and the sample gases.
Refer to the information in the section "General technical specifications (Page 83)".
WARNING
your system. Refer to the information on the nameplate and/or in Chapter "Technical
specifications (Page 83)".