Connecting

7.5 Gas connections

7.5
Gas connections
Possible leakages on gas path
Tighten the threaded joints in accordance with the mounting specifications of the
manufacturer using a suitable open-ended wrench. In the process, make sure that you
counter properly. Otherwise, there is a danger that the gas path will leak.
Danger of poisoning by escaping gas
Components which come into contact with the sample gas must be resistant to it.
Otherwise, there is a danger that the gas path will leak and that sample gas will accumulate
in the analyzer.
Take appropriate precautions to prevent gas from escaping, e.g. by monitoring the flow
outside the analyzer.
Note
Reference gas connection
• The reference gas connection with PVDF screwed gland is only suitable for connecting
• The reference gas connection with ISO female thread is only suitable for connecting
46
WARNING
WARNING
an external reference gas pump, see module nameplate of analyzer module 7MB3020-
xAxxx-xxxx, letter A at 2nd position of order number.
reference gas pressures up to 5000 hPa (abs.), see module nameplate of analyzer
module 7MB3020-xCxxx-xxxx, letter C, D or E at 2nd position of order number.
Operating Instructions, 01/2013, A5E31930403-01