Advertisement

5.2.2

OXYMAT 7

Preparing
1. Isolate the wall-mounted device from power.
2. Open the door. To do this, unscrew the six screws.
3. Remove the equipotential bonding cable from the blank plate of the corresponding slot.
4. Remove the blank plate at the corresponding slot. To do this, unscrew the eight Torx
Wall-mounted device
Operating Instructions, 01/2013, A5E31930403-01
This is no longer required if the analyzer module (AM) is fitted.
CAUTION
Device damage
Before withdrawing the plug connector, loosen the lock by pressing with your index
finger. The contacts or the plug connector could otherwise be damaged.
Figure 5-1
Unlocking the connector for equipotential bonding
screws.
Note
Check the flat gasket on the base of the housing for damage
Replace the gasket if it is damaged or porous.
Installing the analyzer modules and option module
5.2 Installation of the analyzer modules