Connecting

7.1

Safety notes for connection

Cable glands
The cables must have the correct diameter for the cable glands. After installation, make
sure that the cables are positioned securely and that the seals are in the correct position.
Danger due to electrical current
Check that the existing power supply corresponds to that specified on the nameplate of the
device.
Danger due to electrical current
The power supply cable to be connected at the customer location could be damaged by
incorrect fitting in the device.
You should therefore carry out a visual inspection and PE conductor test of the cable in
accordance with the country-specific safety standards, regulations or other applicable
directives before starting up the device. Since this is a cable which lies partially within the
device, you must always additionally apply the production safety standard IEC / EN 61010-
1 or its national version.