Preparing The System For Power Return (If Power Was Unexpectedly Lost While System Was On) - Siemens ADVIA 2400 Maintenance Procedures

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

3. When the power returns, perform the normal startup operation.
Preparing the system for power return (if power was unexpectedly
lost while system was on)
While the electrical power is still off, do the following:
1. Turn off the workstation power switch.
2. At the power panel, set the
When the electric power returns, do the following:
1. Turn on the workstation power switch.
2. When the Startup window opens, turn the
3. Select the
4. At the Startup window, enter a password.
5. Select
6. If possible, repeat the task that you were performing prior to the power failure and
verify that the data was stored.
7. If reagent was dispensed, you must perform a Weekly WASH2 before resuming
operation.
Recovering from an unexpected power outage (after power returns,
when power was unexpectedly lost while system was on)
1. If the Startup window is open, select
operation.
2. Turn off the workstation power and turn the Operate/Standby switch on the analyzer
to
Standby
3. Wait approximately 20 seconds.
4. Perform the normal startup operation and open the Startup window.
5. At the Startup window, enter a password.
6. Select
7. If possible, repeat the task prior to the power failure and verify that the data was
stored.
8. If reagent was dispensed, you must perform a Weekly WASH2 before resuming
operation.
172
Operate/Standby switch
button on the analyzer unit power supply panel.
system reset
, then select
Re-Start
OK
.
Re-Start
, then select
OK
to Standby.
Operate/Standby switch
.
and perform the normal shutdown
Shutdown
.
to Operate.
ADVIA 2400 Operator's Guide