Every 4 months maintenance

Cleaning the ancillary reagent bottle filters
Materials required:
• Filters (REF 08602474, PN 073-0033-01)
order one set per bottle
Time: 20 minutes
Analyzer mode: OFF
Use this procedure to clean the filters in the following bottles
• RRV (reaction) bath oil
• Diluent
• Cuvette wash
• Conditioner
1. Open the filter cap at the front of each bottle and pull up the filter line.
2. Unfasten the connector at the end of the line.
NOTE
If any of the filters are ripped or damaged, replace them with new filters.
3. Remove the filter and inspect it for particles or dirt.
4. If dirty, clean the filters:
a. Place the filters in a beaker filled with a fresh 10% solution of water and
b. After 30 minutes, remove the filters.
c. Rinse them in deionized water.
d. Replace them into their respective holders.
5. Fasten the connector.
6. Using a pad soaked in 5% bleach solution, clean the outside surfaces of the filter
holders and hoses.
7. Insert the filter hoses into the bottles, then fasten the caps.
8. At the Operation Panel, select
9. At the Prime Set dialog box, select
10. Type
Maintenance
household bleach.
10
or more for the number of times in all fields, then select
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
.
Prime
.
Prime 2
Execute
.
145