Replenishing the RRV (reaction) bath oil bottle

Materials required:
• RRV (reaction) bath oil
(REF 09323099, PN B01-4180-01)
Time: 10 minutes
Analyzer mode: READY
Figure 5-33. RRV bath oil bottle
CAUTION
Do
not
do not mix.
1. Unscrew the filter cap (
tube with the filter.
2. Disconnect the RRV bath oil bottle level sensor connector (
connector counter-clockwise and pull it out.
CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
3. Remove the RRV bath oil bottle (
Ensure that level-sensor connector (
158
attempt to clean the RRV bath oil bottle (
3
) at the front of the RRV bath oil bottle (
CAUTION
to return the probes back to home (over the
Initialize
,
, and then
Prime
PRIME 2
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1
ISE buffer bottle
2
RRV (Reaction) bath oil bottle
3
Isotonic saline diluent bottle
4
Cuvette detergent bottle
5
Cell conditioner bottle
1
) with water; RRV bath oil and water
).
1
2
) does not get wet, to avoid damaging it.
to ensure proper
Execute
1
), then pull up the
), then turn the
2
ADVIA 2400 Operator's Guide