Daily maintenance

Inspecting and cleaning the probes
Materials required:
• Phillips screwdriver
• Alcohol prep pad or lint-free towels and
5% bleach solution
Time: 10 minutes
Analyzer mode: READY
1. Visually inspect each probe daily.
Figure 5-1 Location of probes
• Replace any clogged probes. See
• Clean all probes daily (proceed to step 2).
TIP
Perform the Shutdown Wash and Weekly Wash as scheduled, to prevent the probes
from clogging.
2. Clean each probe using one of the 3 methods described below:
Cleaning the probes using automatic advance probe motion
1. At the Menu Panel, select
2. At the Manual Operation window, double-select the
move as follows:
Probe
dilution probe
sample probe
Reagent probe 1
Reagent probe 2
3. Select
location, then select
110
Guide.
Maint
Code
3.DPPLR
16.SPPLR
37.RPPLR-1
49.RPPLR-2
the number of times necessary to move the probe to the accessible
Move
to close the probe window.
Exit
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
Replacing the Probes
, then select
Manual Operation
code
ADVIA 2400 Operator's Guide
1 Dilution Probe (DPP)
2 Sample Probe (SPP)
3 Reagent Probe 1
(RPP1)
4 Reagent Probe 2
(RPP2)
in the online Operator's
.
for the probe you want to